- Beaker[1]
- Buret[2]
- Laboratory Bottle[8]
- Laboratory Centrifuge[10]
- Laboratory Heating Equipments[1]
- Other Lab Supplies[10]
- Other Chemicals[8]
- Mixing Equipment[7]
- Separation Equipment[1]
- Gas Analyzers[9]
- Other Analysis Instruments[3]
- PH Meters[6]
- Instrument Parts & Accessories[1]
- Microscopes[10]
- Other Optics Instruments[10]
- Refractometers[5]
- Spectrometers[6]
- Other Measuring & Analysing Instruments[2]
- Testing Equipment[1]
- Paper & Paperboard[2]
- Contact Person : Ms. Huang Lijuan
- Company Name : Guangzhou Huarui Chemical Instrument Co., Ltd.
- Tel : 86-020-81079775
- Fax : 86-020-81868066
- Address : Guangdong,Guangzhou,Rm206, 2F No.48, Hailian Road, Haizhu District, Guangzhou City
- Country/Region : China
- Zip : 510000
Related Product Searches:Gas Chromatograph GC102AF-AT,High Quality,Gas Chromatograph, Gas Chromatograph,GC102AF/GC102AT Gas Chromatograph
Gas Chromatograph
Introduction:
GC102AF/AT Gas Chromatograph
1. Computer aided temperature control with high accuracy (less than ±0.1°C)
2. All control and application parameters, including operation parameters of the detector, can be set up through keyboard. Functions, such as self-diagnosis, switch-off protection, document storage and transaction, limit temperature setup thermal trip valve setup, and FID auto-ignition are incorporated inside of the instrument.
3. Large-screen LCD display, supporting Chinese characters, can accurately and clearly present all temperature control setup values, actual values, TCD working electricity supply, sensitivity of FID magnifier, and etc.
4. The instrument supports packed column i.e. on-column sample injection. It is compatible with such devices as instantaneous vaporizing sample injection, gas sample injection with gas sample injection valves, and 0.53mm large diameter capillary column direct sampler.
5. Open-end computer system can apply RS-232 interface and FJ-2000 chromatogram workstation to achieve dual path communication control and data processing.
6. Large capacity column (300mm×280mm×270mm) is easy to be installed with packed column and large diameter capillary column.
Performance Indicators:
Temperature Control | |
Control Areas | gas chromatogram column, sampler, detector |
Control Range | 15°C-399°C of room temperature (increment by 1°C) |
Control Accuracy | less than ±0.1°C (measured at 200°C) |
Flame Ionization Detector (FID)( GC102AF) | |
Detection Limit | Mt≤1×10-10g/s(sample: C16) |
Drift | ≤2×10-12A/h |
Thermal Conductivity Detector (TCD) (GC102AT ) | |
Sensitivity | S≥1500mV·ml/mg(carrier gas: H2; sample: n-Hexadecane) |
Baseline Noise | ≤20µV |
Baseline Drift | ≤50µV/h |
GC102AT/GC102AF:
Optional Accessories | 1. CDP-4A Data Processor 2. Software with I/O Port |
Shipping Package Dimension | 950×570×770mm |
Net Weight | 45kg |
Gross Weight | 65kg |
Gas Chromatograph GC102AF-AT